Auxiliary Recipes
Bias Image
- Mode:
Bias Image
- Recipe class:
- Input class:
BiasRecipeInput
- Result class:
BiasRecipeResult
The actions to calibrate the zero (pedestal) level of the detector plus associated control electronic by taken images with null integration time.
Requeriments
Name |
Type |
Default |
Meaning |
---|---|---|---|
|
Product |
NA |
Master BPM frame |
Procedure
The frames in the observed block are stacked together using the median of them as the final result. The variance of the result frame is computed using two different methods. The first method computes the variance across the pixels in the different frames stacked. The second method computes the variance en each channel in the result frame.
Products
Name |
Type |
---|---|
|
|
|
|
Dark Current Image
- Mode:
Dark Current Image
- Recipe class:
- Input class:
DarkRecipeInput
- Result class:
DarkRecipeResult
The actions to measure the variation of the intrinsic signal of the system by taken images under zero illumination condition and long integration time.
Requeriments
Name |
Type |
Default |
Meaning |
---|---|---|---|
|
Product |
NA |
Master BPM frame |
|
Product |
NA |
Master Bias frame |
Procedure
The frames in the observed block are subtracted from the master bias and then they are stacked together, using the median of them as the final result. The variance of the result frame is computed using two different methods. The first method computes the variance across the pixels in the different frames stacked. The second method computes the variance en each channel in the result frame.
Products
Name |
Type |
---|---|
|
|
|
|
Intensity Flat-Field
- Mode:
Intensity Flat-Field
- Recipe class:
- Input class:
IntensityFlatRecipeInput
- Result class:
IntensityFlatRecipeResult
The required actions to set the TS and EMIR at the configuration from which sky and/or artificial illumination flat field data acquisition can proceed and take data.
Requeriments
Name |
Type |
Default |
Meaning |
---|---|---|---|
|
Product |
NA |
Master BPM frame |
|
Product |
NA |
Master Bias frame |
|
Product |
NA |
Master Dark frame |
|
Product |
[1.0, 0.0] |
Master non-linearity calibration |
Procedure
The frames in the observed block are subtracted from the master bias and the master dark. The frames are corrected from non-linearity.
The frames with lamps-on and with lamps-off are stacked using the median, and then the combined lamps-off frame is subtracted from the lamps-on frame. The result is the subtracted frame, scaled to have a mean value of 1.
Products
Name |
Type |
---|---|
|
|